Kestra Announces Publication of ACE-PAS, the Largest Prospective Real-World Wearable Defibrillator Study, in the Journal of Cardiovascular Electrophysiology

$KMTSPress releaseNov 16, 2026, 6:00 AM ETRead the release

Kestra announced peer-reviewed publication of ACE-PAS primary results in the Journal of Cardiovascular Electrophysiology — the largest prospective real-world wearable defibrillator study to date, enrolling 21,612 patients across 35 U.S. states. The ASSURE WCD met both primary endpoints: 100% all-shock conversion success across 303 analyzable VT/VF events and an inappropriate-shock rate of 0.0065 per patient-month that beat the prespecified safety goal. Median daily wear was 23.1 hours.

Key figures

Drug name
ASSURE WCD
Phase of trial
Post-approval registry study (ACE-PAS)
States covered
35
Patients enrolled
21,612
Pct use beyond 90 days
30%
Median daily wear hours
23.1
Shock events analyzable
303
Shock conversion success
100%
Vt vf 90day risk post mi
3.2%
Pct no false positive alarms
94%
Pct patients with vt vf event
2.6%
Vt vf 90day risk ischemic cmp
2.8%
Vt vf 90day risk nonischemic cmp
2.5%
Inappropriate shock rate per patient month
0.0065
Projected inappropriate shock rate reduction
nearly 70% (0.0065 to 0.0020 per patient-month) with enhanced detection algorithm